LBS180
3-hour Steam Sterilization Biological Indicator
• 3 hour results • Rapid Fluorescence and exact results • Reduced incubation time accuracy > 97% • Dramatic color change • Kit includes both BI ampoules and culture media tubes
• 3 hour results • Rapid Fluorescence and exact results • Reduced incubation time accuracy > 97% • Dramatic color change • Kit includes both BI ampoules and culture media tubes
Brand | LISTER | |
Process | Steam | |
Microorganism | Geobacillus Stearothermophilus | |
Population | 10^6 Spores/carrier | |
Read-Out Time | 3 hour | |
Resistance | 121℃ D value / m | 1.5 to 4 min |
121℃ Survival / m | 5 to 15 min | |
121℃ Kill / m | 18 to 40 min | |
132℃ D value / m | 0.15 to 0.7 min | |
134℃ D value / m | 0.1 to 0.6min | |
Z value | 8 to 30℃ | |
RIT(reduced incubation time accuracy) | ≥97% | |
Test condition | 121℃; 132℃; 134℃ Saturated steam with pre-vacuum |
Label color change | Before: pink; After: black; |
Color of culture medium | Positive:yellow; Negative:purple; |
Quality of culture medium | 0.5-0.6ml |
Weight of BI | 2.3-2.4g/pcs |
Period of validity | 24 months |
Dimensions | 10.9 mm (cap diameter) |
8.6 mm (pipe diameter) | |
57 mm (height) | |
Regulations | ISO 11138-1:2017 |
ISO 11138-2:2017 | |
ISO 11138-8:2021 | |
ISO 11140-1:2014 | |
Package | 50 pcs/box |
Rapid Biological Indicator For Steam sterilization processes –LBS180
• 3 hour results
• Rapid fluorescence and exact results
• Reduced incubation time accuracy ≥97%
• Dramatic color change
• Kit includes both BI ampoules and culture media tubes
The LBS180 is utilized to monitor sterilization efficacy of gravity displacement or vacuum-assisted steam processes at 121°C to 135°C.
Related Document Downloads
Product Brochure of Biological Indicators.pdf
Technical Date Sheet LBS180.pdf
Self-Contained Biological Indicators consist of: ✅ A plastic pipe and cap. The cap with a hole is placed on top of the pipeto allow the sterilizing agent to penetrate into the indicator. Between pipe and cap there is a filter paper which is used to avoid pollution. ✅ A crushable media ampule which contains modified culture mediumwith a pH indicator and fluorogenic substrates. The modified culture mediumwill transition from the initial Purple color to Yellow in the presence of bacterial growth. ✅ An inoculated carrier of Geobacillus stearothermop hilus ATCC 7953 is located at the bottom closed end of the pipe. ✅ A barrier is positioned like a plug between the carrier and the ampule. ✅ The LBS180 Biological Indicator, when used with an automatic reader(incubation temperature 56-60°C), provides a fluorescence test result in a short period(180 minutes) with an accuracy of ≥97%. As an option the user can continue the incubation for 48 hours in order to confirm visually the stability or color change of the culture medium. ✅ The label on the outside of the plastic pipe contains a Type 1 chemical indicator strip. |